Page 11-6 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Control Test
Use the Control Test to verify that the ventilator buttons and the Set Values knob are working
correctly.
To run the Control Test:
1) Push the Select button while CONTROL is displayed.
2) SELECT is displayed in the display window.
Although the LTV
®
1000 front panel is shown below, the test is applicable to all LTV
®
Series
ventilators.