Page 10-8 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Variable Rise Time
Use the Variable Rise Time option to select the rise time profile for Pressure Control and Pressure
Support breaths. The rise time profiles are numbered 1 through 9 where 1 is the fastest rise time and
9 is the slowest rise time. Starting with the fastest rise time, each time is 33% longer than the
previous one.
To modify the Rise Time Profile:
1) Push the Select button while RISE TIME is displayed.
2) PROFILE x is displayed.
3) Turn the Set Value knob until the desired Rise Time Profile is displayed.
4) Push the Select button.
Range: 1 to 9, where 1 is the fastest and 9 is the slowest