Series ventilator Page 10-3
p/n 10664, Rev. Y
Alarm Operations
Use the Alarm Operations menu to set up alarm conditions that are not available directly from front
panel controls. The menu is set up as follows:
ALARM OP
ALARM VOL
APNEA INT
HP DELAY
LPP ALARM
HIGH f
HIGH PEEP
PNT ASSIST
EXIT
Alarm Volume
Use this menu item to set the loudness of the audible alarm.
To modify the Alarm Volume:
1) Push the Select button while ALARM VOL is displayed.
2) VOL xx dBA is displayed.
3) Turn the Set Value knob until the desired setting is displayed.
4) Push the Select button.
Range: 60 - 85 dBA
NOTE
Fixed Volume Alarms – The volume of the Battery Empty alarm cannot be lowered. For patient
safety, this alarm always sounds at full volume.
If the battery depletes to point that the ventilator goes INOP, the Vent INOP audible alarm sounds at
maximum volume for a minimum of 5 minutes. The Vent INOP audible alarm is capacitor driven and
the volume is therefore not user modifiable.
REMARQUE
Alarmes à volume fixe - Le volume de l’alarme de batterie faible ne peut être réduit. Pour la sécurité
des patients, le volume de cette alarme est toujours fort.
Si la batterie est suffisamment déchargée pour que le voyant INOP du ventilateur s’active, l’alarme
sonore Vent INOP opérera à son maximum de volume pendant 5 minutes au moins. L’alarme sonore
Vent INOP est contrôlée par un condensateur et son volume ne peut donc pas être modifié par
l’utilisateur..