Page 6-2 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Breath Rate
Use the Breath Rate control to establish the minimum rate of machine or assist breaths that the
ventilator will deliver per minute.
To set the Breath Rate:
1) Push the Breath Rate button.
2) Change the setting using the Set Value knob.
Range: “- -”, 1 - 80 bpm
NOTE
When SIMV/CPAP is selected, the ventilator will be in SIMV or CPAP mode, depending on the Breath
Rate setting.
If Breath Rate is set to dashes “- -”, the ventilator will be operating in CPAP mode.
If Breath Rate is set to any other value, the ventilator will be operating in SIMV mode.
REMARQUE
Lorsque le mode SIMV / CPAP est sélectionné, le ventilateur sera en mode SIMV ou CPAP, selon le
réglage du débit respiratoire.
Si le débit respiratoire est réglé sur Traits « - - », le ventilateur fonctionnera en mode CPAP.
Si le débit respiratoire est réglé sur toute autre valeur, le ventilateur fonctionnera en mode SIMV.