EasyManua.ls Logo

Pulmonetic Systems LTV 950 - Control Lock

Default Icon
277 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual LTV
®
Series ventilator Page 6-3
p/n 10664, Rev. Y
Control Lock
The LTV
®
Series ventilator front panel controls may be locked so that settings are not accidentally
changed. Two different levels of difficulty can be set for control unlocking: Easy and Hard. Easy
unlocking is the default and this setting is changed using the Extended Features menus
14
. For more
information on using the Control Lock, see Chapter 5 - Control Locking.
To turn the Control Lock on:
1) Push the Control Lock button.
The Control Lock LED is on whenever the front panel controls are locked.
To turn the Control Lock off with Easy unlocking:
1) Push the Control Lock button.
To turn the Control Lock off with Hard unlocking:
1) Push and hold the Control Lock button for 3 seconds.
These controls are not affected by the control lock and operate even when the control lock is on:
Manual Breath, Silence / Reset, Select.
14
See Chapter 10 - Extended Features, Control Unlock for more information.