Operator’s Manual LTV
®
Series ventilator Page 9-3
p/n 10664, Rev. Y
BAT EMPTY
When the ventilator is operating on internal battery power and the battery charge level falls below the
empty threshold, the BAT EMPTY alarm is generated. For patient safety, this alarm can only be
silenced once for 30 seconds and cannot be silenced preemptively. After the 30 seconds has
elapsed, the alarm cannot be silenced again and will continue to sound and display until an alternate
power source is provided or the battery is depleted and the ventilator goes INOP. This alarm will
always sound at maximum volume.
WARNING
BAT EMPTY alarm - A BAT EMPTY alarm indicates the internal battery is almost depleted. Connect
the ventilator to an external power source immediately.
AVERTISSEMENT
Alarme BAT EMPTY - Une alarme BAT EMPTY indique que la pile interne est pratiquement à plat.
Branchez immédiatement le ventilateur à une source d'alimentation externe.
When a BAT EMPTY alarm occurs:
• The Battery Level LED displays red.
• The BAT EMPTY message is displayed.
• The audible alarm is sounded.
NOTE
Volume of the BAT EMPTY alarm cannot be lowered. For patient safety, this alarm will always sound
at full volume
REMARQUE
Le volume de l’alarme batterie faible ne peut être réduit. Pour la sécurité des patients, le volume de
l’émission de cette alarme est toujours fort.
To temporarily silence the BAT EMPTY alarm:
1) Push the Silence / Reset button once to silence the audible alarm for 30 seconds only.
The BAT EMPTY alarm can only be silenced once and cannot be reset until the battery is recharged
or external power is applied.