Page E-6 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Event Codes by Event Name
Event Name Code Event Associated Alarm
AD MMTCH 33 ADC mismatch HW FAULT
AD MTCH0 35 ADC mismatch cleared HW FAULT
AD MTCH1 34 ADC mismatch occurred HW FAULT
ADC1 VAL 65 AD mismatch primary channel fault value HW FAULT
APNEA 0 06 Apnea mode exited APNEA
APNEA 1 05 Apnea mode entered APNEA
BATLOW0 14 Internal battery low exited BAT LOW
BATLOW1 13 Internal battery low occurred BAT LOW
BATMPT0 12 Internal battery empty exited BAT EMPTY
BATMPT1 11 Internal battery empty occurred BAT EMPTY
CLR BREC 88
Reclaims all incorrectly recognized bad
EEPROM records
N/A
CLR CAL 60 Calibration records cleared N/A
CLR CTRL 52 Control settings cleared N/A
CLR EVNT 51 Event log cleared N/A
CRC 41 Memory CRC check failed RESET
DEFAULTS 25 Defaults, or Set Defaults occurred
DEFAULTS / DEFAULTS,
SET
DISC 0 10 Circuit disconnect exited DISC/SENSE
EEPROM 40 EEPROM degraded HW FAULT
EXT LOW0 18 External power low exited POWER LOW
EXT LOW1 17 External power low occurred POWER LOW
EXT LST0 16 External power lost exited POWER LOST
EXT LST1 15 External power lost occurred POWER LOST
FAN FLT0 28 Fan fault exited HW FAULT
FAN FLT1 27 Fan fault occurred HW FAULT
FLUSH ER 68
A problem is detected writing data to the
EEPROM during system shutdown.
HW FAULT
HI PEEP0 76 High PEEP alarm recovered HIGH PEEP
HI PEEP1 75 High PEEP alarm occurred HIGH PEEP
HI PRES0 43 High pressure exited HIGH PRES
HI PRES1 42 High pressure occurred HIGH PRES
HIGH DIS 08 High side disconnect DISC/SENSE
HIGH f0 74 High breath rate alarm recovered HIGH f
HIGH f1 73 High breath rate alarm occurred HIGH f
HOME ER0 39 Stepper motor home failure exited HW FAULT
HOME ER1 38 Stepper motor home failure occurred HW FAULT
HOUR MTR 03 Set hour meter None
INTRRPT1 31 Spurious interrupt occurred ms RESET 1
INTRRPT2 32 Spurious interrupt occurred ls RESET 1
LN VENT1 67 Shutdown for other than pressing On/Standby RESET