Operator’s Manual LTV
®
Series ventilator Page E-7
p/n 10664, Rev. Y
Event Name Code Event Associated Alarm
LO PRES0 50 Low peak pressure exited LOW PRES
LO PRES1 49 Low peak pressure occurred LOW PRES
LOW DIS 09 Low side disconnect DISC/SENSE
LOW VE 0 48 Low minute volume exited LOW MIN VOL
LOW VE 1 47 Low minute volume occurred LOW MIN VOL
NO CAL 26 No calibration data found NO CAL DATA
O2 HI 0 24 O
2
pressure high exited HIGH O2 PRES
O2 HI 1 23 O
2
pressure high occurred HIGH O2 PRES
O2 LOW 0 22 O
2
pressure low exited LOW O2 PRES
O2 LOW 1 21 O
2
pressure low occurred LOW O2 PRES
POST 57 POST failure RESET
RAC ERR0 70
Recovery from problem detected with primary
and/or redundant audible alarm circuitry
HW FAULT
RAC ERR1 69
Problem detected with primary and/or
redundant audible alarm circuitry
HW FAULT
RUNAWAY 58 Code runaway detected RESET
SET DATE 53 Date set N/A
SET TIME 54 Time set N/A
SNDRERR0 72 Recovery from alarm sounder error HW FAULT
SNDRERR1 71 Alarm sounder error HW FAULT
STACK 56 Stack overflow detected RESET
SYNC ER1 36 Stepper motor lost sync occurred HW FAULT
SYNCER0 37 Stepper motor lost sync exited HW FAULT
TBN ESTP 46 Turbine emergency stop occurred HIGH PRES
TBN HSTP 66 Turbine Hold Stop occurred HIGH PRES
TBN ISTP 44 Turbine immediate stop occurred HIGH PRES
TBN ZERO 45 Turbine set to zero flow occurred HIGH PRES
VENT 0 02 Power off None
VENT 1 01 Power on None
VENT CHK 04 Set vent check
Entered VENT CHECK
mode
WDOG TST 59 Watchdog test run Inop
XDC FLT0 20 XDCR fault exited XDCR FAULT
XDC FLT1 19 XDCR fault occurred XDCR FAULT
XDCR AIR 64 Airway pressure transducer fault XDC FLT1
XDCR BI 63
Differential pressure transducer - Bi-
directional channel fault
XDC FLT1
XDCR NAR 61
Differential pressure transducer - Narrow
channel fault
XDC FLT1
XDCR WID 62
Differential pressure transducer - Wide
channel fault
XDC FLT1