EasyManua.ls Logo

Pulmonetic Systems LTV 950 - Display Test

Default Icon
277 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 11-4 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Display Test
Use the Display Test to verify that the ventilator displays are working correctly.
To run the Display Test:
1) Push the Select button while DISPLAY is displayed.
2) All segments of the 7-segment control displays, all dots of the dot-matrix window displays and all
LEDs are illuminated.
Although the LTV
®
1000 front panel is shown below, the test is applicable to all LTV
®
Series
ventilators.
NOTE
The display states for the External Power, Vent Inop, and Charge Status LEDs are not tested in the
Display Test.
The External Power and Charge Status LEDs are tested and verified when the AC adapter is
connected to the ventilator (see page 11-2).
The Vent Inop LED is tested and verified during the Vent Inop Alarm Test (see page 11-11).
REMARQUE
Ce test d’affichage ne comprend pas l’état d’affichage des DEL External Power, Vent Inop, et
Charge Status.
Les DEL External Power et Charge Status sont testés et vérifiés lorsqu’on branche l’adapteur
CA au ventilateur (voir page 11-2).
Le DEL Vent Inop est testé et vérifié en même temps que l’avertisseur du Vent Inop (voir page
11-11).

Table of Contents

Other manuals for Pulmonetic Systems LTV 950

Related product manuals