NO CAL DATA, NO CAL Monitor Display
When the ventilator detects invalid or missing calibration records on power up, the NO CAL DATA
alarm is generated. When this happens, default calibration values are used, and although the
ventilator will continue to operate, the accuracy of volumes and pressures may be reduced.
A NO CAL message is posted in place of affected monitored values when the ventilator is operating
without valid transducer calibration data.
When the NO CAL DATA alarm occurs:
The NO CAL DATA message is flashed in the display window.
The audible alarm is sounded.
The ventilator continues to operate.
Default transducer data is used.
Vte, PIP, MAP, PEEP, and VE monitored values are displayed as NO CAL.
To clear the NO CAL DATA alarm:
1) Push the Silence / Reset button twice
. This will clear the alarm and the ventilator will continue to
operate; however, the NO CAL message will still be displayed in place of affected monitored
values.
2) Take the unit out of service and perform the Calibration procedure
45
.
To clear the NO CAL message:
1) Take the unit out of service and perform the Calibration procedure
45
.
WARNING
NO CAL Condition - Operation of the LTV
®
Series ventilator under a NO CAL condition may result in
inaccurate pressure and volume measurements. Should this condition occur, disconnect the patient
from the ventilator, provide an alternative method of ventilation and immediately contact a certified
Pulmonetic Systems’ service technician or Pulmonetic Systems.
AVERTISSEMENT
Condition NO CAL - L'opération continue du ventilateur de la série LTV
®
sous condition NO CAL
peut résulter en mesures de pression et de volume erronées. Si cette condition se présente, le
ventilateur doit être retiré du service, et vous devez immédiatement contacter votre technicien de
service certifié de Pulmonetic Systems ou Pulmonetic Systems.
45
For more information on the Calibration procedure, see the LTV® Series ventilator Service Manual.