Page 11-8 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Leak Test
Use the Leak Test to test the patient circuit for leaks.
To run the Leak Test:
1) Attach all patient circuit accessories (such as water traps, heated circuits and humidifiers) to the
patient circuit.
2) Connect the patient circuit to the LTV
®
Series ventilator.
3) With a clean, gloved hand or 4”X4” gauze pad, occlude the proximal end of the patient circuit.
4) Push the Select button while LEAK is displayed.
NOTE
The Leak Test cannot be run until the ventilator has been running for 60 seconds. If you attempt to
run the leak test before the warm-up period has completed, a WARMUP xx message will be
displayed. When the warm-up period is complete, the Leak Test menu item is redisplayed.
REMARQUE
Le test de fuites ne peut s'exécuter tant que le ventilateur n'a pas fonctionné durant 60 secondes.
Si vous tentez d'exécuter un test de fuites avant que la période de réchauffement ne soit
complétée, un message «WARMUP xx» sera affiché. Lorsque la période de réchauffement est
complétée, les éléments du menu Test de fuites sont de nouveau affichés.