REMOVE PTNT
When the ventilator is powered up in the Ventilator Checkout or Ventilator Maintenance modes, the
REMOVE PTNT alarm is generated to remind you to remove the patient from the ventilator before
proceeding. Use the Ventilator Checkout mode to check for correct operation of the displays and
controls and to check the patient circuit for leaks. Ventilator Maintenance mode is used by technical
personnel to perform maintenance or calibration.
WARNING
Ventilator Checkout and Maintenance Modes - The LTV
®
Series ventilator does not deliver gas
during the Ventilator Checkout mode (VENT CHECK) or Ventilator Maintenance mode (VENT
MTNCE) and should not be used to ventilate a patient during these tests.
AVERTISSEMENT
Modes Vérification et Entretien du ventilateur - Le ventilateur de la série LTV
®
ne transmet pas le
mélange de gaz en mode Vérification du ventilateur (VENT CHECK) ou en mode Entretien du
ventilateur (VENT MTNCE), il ne devrait donc pas être utilisé pour ventiler un patient durant
l'exécution de ces tests.
When you enter Ventilator Checkout mode or Ventilator Maintenance mode, a REMOVE PTNT
alarm occurs:
• The REMOVE PTNT message is displayed.
• The audible alarm is sounded.
To reset the REMOVE PTNT alarm:
1) Push the Silence / Reset button twice
.