Page 6-20 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
On / Standby
This button switches the LTV
®
Series ventilator between Standby and On.
When the ventilator is on, the On / Standby LED will be on. The ventilator will operate on external
power if it is available, or internal battery, if there is no external power or the external power source is
depleted. The internal battery will be charged from the external power source while the ventilator is
operating on external power.
When the ventilator is in Standby, the On / Standby LED will be off, however, the internal battery will
continue to charge.
To turn the ventilator on from the Standby state:
1) Push the On / Standby button.
To put the ventilator into Standby:
1) Push and hold the On / Standby button for 3 seconds.
2) An Inop alarm will occur. To cancel the Inop alarm, push the Silence/Reset button.
• For ventilators with an audio sound symbol (
) on the back panel label, verify a confirming
audible chirp occurs after the alarm is silenced.
3) The VENT INOP LED will remain lit for a minimum of 5 minutes.