Page 10-16 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Software Version
Use the Software Version option to verify the software version installed in the ventilator. The software
version number is displayed as: VER xx.xx X
Usage Meter
Use the Usage Meter to view the time the ventilator has been in use. It is updated every 1/10
th
hour
up to 139,000.0 hours and is displayed as: USAGE xxxxxx.x
Communications Setting
The ventilator may be connected to a printer, a graphics monitor, or a modem, or may be set up to
output system diagnostic data. Use the Communications Setting option
51
to select the
communications protocol for data transmission.
• For LTM™ Compatible LTV
®
900, 950 and 1000 Ventilators
52
, use the MONITOR setting to
communicate with an LTM Graphics Monitor. If the MONITOR setting is not available, the LTV
®
ventilator being used requires upgrades before it can support the LTM Graphics Monitor. Contact
your Pulmonetic Systems Service Representative for additional information.
To modify the Communications Setting:
1) Push the Select button while COM SETTING is displayed.
2) DATA or the currently selected protocol is displayed.
3) Turn the Set Value knob until the desired protocol is displayed.
4) Push the Select button.
Options: DATA, MONITOR, PRINTER, and MODEM
51
Only DATA and MONITOR (for LTM Graphic Monitors compatible LTV
®
Ventilators) are available at this time.
52
LTM Graphics Monitor compatibility can be verified by pushing the Select button when the LTV Model
Number is displayed in the Extended Features menu. The message LTM will be displayed if the ventilator was
originally manufactured or upgraded by Pulmonetic Systems to accommodate the LTM Graphics Monitor.