Operator’s Manual LTV
®
Series ventilator Page 9-9
p/n 10664, Rev. Y
DEFAULTS SET
The DEFAULTS SET alarm is generated when the LTV
®
Ventilator is first powered up after the SET
DEFAULTS
40
option has been used to reset all controls and extended features settings to their
factory-set default values
41
.
Language, Time/Date Format and Com settings are not reset to default values when the SET
DEFAULTS option is used in Extended Features.
When a DEFAULTS SET alarm is generated:
The DEFAULTS SET message is flashed in the display window.
The audible alarm is sounded.
To reset the DEFAULTS SET alarm:
1) Push the Silence / Reset button twice.
2) Select and return the control(s) and extended features settings to the desired settings.
NOTE
Be sure to check all Controls, Alarms and Extended Features options and return them to the desired
settings.
REMARQUE
Assurez-vous de procéder à la vérification de toutes les options de contrôles, d'alarmes et de
caractéristiques étendues, et de les retourner aux réglages souhaités.
40
See Chapter 10 - Extended Features, Set Defaults for additional information.
41
See Chapter 9 - Ventilator Alarms, DEFAULTS for factory-set default values.