Operator’s Manual LTV
®
Series ventilator Page 9-7
p/n 10664, Rev. Y
DEFAULTS
All controls and extended features on the LTV
®
Series ventilator have factory-set default values.
When the operator makes changes to the controls or extended features settings, the ventilator stores
the new settings in non-volatile memory
38
. During POST, the ventilator checks the stored settings. If
the ventilator detects an invalid stored setting, the DEFAULTS alarm occurs and the affected settings
are set to the default values.
When a DEFAULTS alarm is generated:
• An audible alarm is sounded.
• The DEFAULTS message is flashed in the display window.
• All affected controls or features are set to their default values.
To reset the DEFAULTS alarm:
1) Push the Silence / Reset button twice.
2) Select and return the control(s) or features to the desired settings.
NOTE
Be sure to check all Controls, Alarms and Extended Features options and return them to the desired
settings.
Repeated occurrences of the DEFAULTS alarm may indicate a problem with the ventilator’s non-
volatile memory. Please immediately contact a certified Pulmonetic Systems’ service technician.
Control values are re-set to default values each time the ventilator is turned on, only if an invalid
stored setting is detected during POST.
REMARQUE
Assurez-vous de procéder à la vérification de toutes les options de contrôles, d'alarmes et de
caractéristiques étendues, et de les retourner aux réglages souhaités.
L'occurrence répétitive des alarmes PAR DÉFAUT peut indiquer un problème avec la mémoire non
volatile du ventilateur. Veuillez contacter immédiatement un technicien de service certifié de
Pulmonetic Systems.
Les valeurs de contrôle sont rétablies à leurs valeurs par défaut chaque fois que le ventilateur est
allumé, seulement si un paramètre en mémoire non valable est détecté au moment du diagnostic
automatique de mise sous tension.
38
Non-volatile memory is memory that is not erased when the ventilator is turned off or disconnected.