Page 9-18 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
LOW MIN VOL
When the exhaled minute volume (VE) is less than the Low Minute Volume setting, the LOW MIN
VOL alarm is generated.
To prevent nuisance alarms, the LOW MIN VOL alarm is suspended for the first 20 seconds of
ventilator operation after power up and passing the Power On Self Tests.
When a LOW MIN VOL alarm occurs:
The LOW MIN VOL message is flashed in the display window.
The Low Minute Volume Control display is flashed.
The audible alarm is sounded.
To reset the LOW MIN VOL alarm:
1) Push the Silence / Reset button twice
.
This alarm is not available in NPPV mode.
WARNING
Low Minute Volume Control Settings - The Low Minute Volume control should be set to its highest
clinically appropriate value. If there is a clinical need to set the Low Minute Volume alarm to lower
values or off (“- - -“), perform a clinical assessment to determine if an alternative monitor (i.e. a Pulse
Oxymeter with an audible alarm, or a Cardio Respiratory Monitor) should be used.
AVERTISSEMENT
Réglages du contrôle de volume bas par minute - Le contrôle du volume bas par minute doit être
ajusté à la plus haute valeur clinique appropriée. Si l’alarme de volume bas par minute doit être
ajustée à des valeurs inférieures ou mise à l’arrêt (“- - -“) pour satisfaire aux besoins cliniques,
effectuer une évaluation clinique afin de déterminer si l’utilisation d’un autre moniteur (c.-à-d.,
sphygmo-oxymètre muni d’une alarme sonore ou un moniteur cardio-respiratoire) s’avère pertinente.