Page 13-4 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Cleaning or Replacing the O2 Inlet Filter
The O
2
filter should be cleaned or replaced when it becomes soiled. Failure to do this can affect
ventilator performance.
CAUTION
Oxygen Supply Contamination - The accuracy of the oxygen delivery capabilities of LTV
®
ventilators can be compromised by foreign debris contamination in the oxygen supply system. To
reduce the risk of airborne contaminants entering the ventilator, ensure that any oxygen supply
connected to the ventilator is clean, properly filtered
64
and that the ventilator’s O
2
Inlet Port Cap is
securely installed on the O
2 I
nlet Port whenever the ventilator is not connected to an external oxygen
supply.
ATTENTION
Contamination de la réserve d’oxygène - La précision de la capacité d’alimentation en oxygène
des ventilateurs LTV
®
peut être compromise par la présence de corps étrangers dans le système
d’alimentation en oxygène. Afin de diminuer le risque de présence d’agents contaminants
atmosphériques dans le ventilateur, assurez-vous que la réserve d’oxygène reliée au ventilateur est
propre et filtrée de manière adéquate
64
, et que le bouchon de l’orifice d’alimentation en oxygène est
correctement installé à chaque fois que le ventilateur n’est pas relié à une source d'oxygène externe.
To clean or replace the O
2
Inlet Filter:
1) If a high pressure O
2
source is being used, disconnect the high pressure O
2
hose from the
oxygen block on the left side of the ventilator.
2) If a low pressure O
2
source is being used, disconnect the O
2
line from the barbed oxygen
adapter. Unscrew and remove the barbed adapter from the oxygen block on the left side of the
ventilator.
64
In addition to the existing internal O
2
Inlet filter, P/N 14313, an External, In-Line Oxygen Filter (P/N 14470) is
available from Pulmonetic Systems, Inc.