EasyManua.ls Logo

Pulmonetic Systems LTV 950 - Alarm Test

Default Icon
277 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operator’s Manual LTV
®
Series ventilator Page 11-3
p/n 10664, Rev. Y
Alarm Test
Use the Alarm Test to verify that the audible alarm is working correctly.
To run the Alarm Test:
1) Push the Select button while ALARM is displayed.
2) Verify that the audible alarm is sounded.
If a Patient Assist Call System or Remote Alarm is connected via the ventilator’s Patient
Assist Port, verify the device also activates (audible/visual), as specified by its manufacturer.
3) When the alarm has sounded for at least 2 seconds, push the Select button again.
The audible alarm is silenced and the next menu item is displayed.
4) For ventilators with an audio sound symbol (
) on the back panel label, verify a confirming audible
chirp occurs after the alarm is silenced.
If the Alarm Test fails, see Chapter 15 - Troubleshooting for more information.

Table of Contents

Other manuals for Pulmonetic Systems LTV 950

Related product manuals