Page A-6 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Storage and Operating Conditions (cont.)
Specification
Tolerance
Oxygen Inlet
DISS or NIST Connector Inlet Pressure Range
76
40 to 80 PSIG
± 2 PSIG
Tapered Tubing Connector Inlet Pressure Range 0 to 10 PSIG
± 2 PSIG
Shock and Vibration
The ventilator is designed to withstand shock and vibration in accordance with relevant
requirements set forth in the following standards:
IEC 68-2-27 Shock
IEC 68-2-6 Vibration
IEC 68-2-34 Vibration
MIL-STD-810E Shock, Ground Transport and Helicopter Transport Vibration
Spillage
The ventilator resists fluid spillage when tested in accordance with the relevant standards
specified in IEC 601-1 Clause 44.3.
External Surface Temperature
External surfaces
< 50°C, ambient temperature of 35°C
n/a
Communications
Port Connector Specification
Communications RS232, DB9 connector Protocol Options: Data, Monitor, Printer,
Modem
Patient Assist Call /
Remote Alarm
RJ11-4
Closed contact resistance: ≤ 1 ohm
Equipment Classification
Classification The ventilator is rated as Class II equipment per IEC 601-1 Clause 6.11
Type The ventilator is specified as Type BF equipment per IEC 601-1 Clause 6.11
76
Not applicable on LTV
®
900 and 950