Operator’s Manual LTV
®
Series ventilator Page A-7
p/n 10664, Rev. Y
Power
Feature Range Tolerance / Indicators
Input Voltage 11 to 15 VDC
External Power
AC Adapter Input: 100 to 250 VAC, 50 to 60
Hz
Output: 13 VDC
± 2.5%
Full Power
Voltage 11.5 V ± 2%
Green LED
Low Power
Voltage < 11.0V and 9.5V ± 2%
Amber LED
External Power Off Voltage < 9.5V
± 2%
LED off, switch to
battery
Hysteresis Ventilator shall not resume
external power operation unless
voltage is 11.5V
± 2%
Nominal Current
Draw
Startup: 5.5 amps
Running: 3-4 amps
Nominal Power
Draw
Startup: 66 watts
Running: 36 - 48 watts
Leakage Current Total leakage current to Earth ground for the ventilator with only
approved accessories attached, shall not exceed 500 microAmps
during normal operation, per IEC 601-1.
Total leakage current to Earth ground for the ventilator shall not exceed
one milliAmp when any single fault condition is present, per IEC 601-1.
Ground Resistance Total impedance between the ground contact at the inlet power
connector and any accessible metal part shall not exceed 0.1 ohm, per
IEC 601-1.
Dielectric Strength The ventilator shall be able to survive 1500 volts applied from either
phase of the AC power inlet to Earth ground for a period of one minute,
per IEC 601-1.