Operator’s Manual LTV
®
Series ventilator Page 9-5
p/n 10664, Rev. Y
BAT LOW
When the ventilator is operating on internal battery power and the battery charge level falls below the
low threshold, a BAT LOW alarm is generated.
When a BAT LOW alarm occurs:
• The Battery Level LED is displayed Amber.
• The BAT LOW message is displayed.
• The audible alarm is sounded.
To reset the BAT LOW alarm:
1) Push the Silence / Reset button twice.
WARNING
Battery run time: When the battery reaches the BAT LOW level, the ventilator will only run for
approximately 10 minutes before generating a battery empty alarm (BAT EMPTY). This approximate
time is based on tests using the nominal settings, a new battery and a full 8 hour charge cycle as
specified in Appendix A - Ventilator Specifications. Actual run time may be more or less depending on
ventilator settings, patient demand, and battery age or condition. It is highly recommended that an
alternate power source is connected PRIOR to the ventilator reaching the BAT EMPTY alarm
condition to ensure continuous, uninterrupted patient ventilation
AVERTISSEMENT
Durée d’utilisation de la batterie – Lorsque la batterie atteint le niveau BAT INT BASS, le
ventilateur fonctionne pendant environ 10 minutes avant d’émettre une alarme de batterie faible (BAT
INT VIDE). Cette durée approximative est basée sur des tests avec des paramètres nominaux, une
nouvelle batterie et un cycle de chargement complet de 8 heures, tel que spécifié dans l’Annexe A –
Spécifications du ventilateur. La durée d’utilisation réelle pourrait être supérieure ou inférieure, selon
les paramètres du ventilateur, la demande du patient et l’âge ou l’état de la batterie. Il est fortement
recommandé qu’une source d’alimentation alternative soit connectée AVANT que le ventilateur
n’atteigne l’état d’alarme BAT INT VIDE afin d’assurer une ventilation continue et ininterrompue au
patient