Operator’s Manual LTV
®
Series ventilator Page 14-3
p/n 10664, Rev. Y
Using an External Battery
Optional External Batteries
67
, Cables
68
and Charger
69
are available from Pulmonetic Systems. The
Large External Battery Pack includes a large capacity battery and hard case with a fuse and power
cable and is pre-wired with a locking quick-connector. The Small External Battery Pack includes a
medium capacity battery, soft bag and power cable with fuse and locking quick-connector.
CAUTION
External Battery Pack - The External Battery Pack should only be connected to the LTV
®
Series
ventilators using the Pulmonetic Systems External Battery Cable (PN 10802). This cable is pre-wired
and properly terminated to ensure safe connection of the External Battery Pack to the ventilator.
Release Button - To avoid damaging the ventilator or the
power connector, push the release button on the connector
before removing it from the ventilator power port or the
power port pigtail connector.
ATTENTION
Bloc-piles externe – Le bloc-piles externe ne doit être branché qu’aux ventilateurs de la série LTV
®
à l’aide du câble pour piles externes de Pulmonetic Systems (N
o
pièce 10802). Ce câble est précâblé
et ses terminaisons assurent une connexion sécuritaire entre le bloc-piles externe et le ventilateur.
Bouton de déclenchement – Pour éviter d'endommager le ventilateur ou le connecteur
d'alimentation, appuyer sur le bouton de déclenchement situé sur le connecteur avant de le retirer du
port d'alimentation du ventilateur ou du raccord de queue de cochon du port d'alimentation.
To run the ventilator from an external battery:
1) Connect the battery cable quick-connector to the port on the external battery hard case or the soft
bag.
67
Pulmonetic Systems External Battery P/N 10787 and Case P/N 10790.
68
Pulmonetic Systems External Battery Cable P/N 10802.
69
Pulmonetic Systems External Battery Charger P/N 10801.