Page 9-28 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
HIGH f OFF
The HIGH f OFF message is displayed when;
• The High Breath Rate alarm is turned off by being set to HIGH f OFF and
,
• The High PEEP alarm (HIGH PEEP) is turned on.
This is an informational message only. The message is displayed at power up, when monitored data
is being scrolled automatically, and when no front panel activity is detected for 60 seconds.
To clear the HIGH f OFF message:
1) Push any front panel button or turn the Set Value knob.
LMV LPPS OFF
The LMV LPPS OFF message is displayed when;
• The Low Minute Volume alarm is turned off by being set to dashes and
,
• The LPP alarm is set to VC/PC ONLY. When this setting is selected, the Low Pressure alarm
applies only to Volume Control and Pressure Control breaths.
This is an informational message only. The message is displayed at power up, when monitored data
is being scrolled automatically, and when no front panel activity is detected for 60 seconds.
To clear the LMV LPPS OFF message:
1) Push any front panel button or turn the Set Value knob.
This message is not displayed in NPPV mode.
LMV OFF
The LMV OFF message is displayed when;
• The Low Minute Volume alarm is turned off by being set to dashes and
,
• The LPP alarm is set to ALL BREATHS.
This is an informational message only. The message is displayed at power up, when monitored data
is being scrolled automatically, and when no front panel activity is detected for 60 seconds.
To clear the LMV OFF message:
1) Push any front panel button or turn the Set Value knob.
This message is not displayed in NPPV mode.