Page C-12 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Oxygen Lines – Connection Instructions
WARNING
Disabled Oxygen Inlet Pressure Alarms - When the oxygen blending option is not installed, the
Oxygen Inlet Pressure Alarms are disabled.
AVERTISSEMENT
Alarmes de pression d'entrée de l'oxygène désactivées - Lorsque l'option de mélange d'oxygène
n'est pas activée, les alarmes de pression d'entrée de l'oxygène sont désactivées.
CAUTION
Oxygen Supply Contamination - The accuracy of the oxygen delivery capabilities of LTV
®
ventilators can be compromised by foreign debris contamination in the oxygen supply system. To
reduce the risk of airborne contaminants entering the ventilator, ensure that any oxygen supply
connected to the ventilator is clean, properly filtered
90
and that the ventilator’s O2 Inlet Port Cap is
securely installed on the O2 Inlet Port whenever the ventilator is not connected to an external oxygen
supply.
ATTENTION
Contamination de la réserve d’oxygène — La précision de la capacité d’alimentation en oxygène
des ventilateurs LTV
®
peut être compromise par la présence de corps étrangers dans le système
d’alimentation en oxygène. Afin de diminuer le risque de présence d’agents contaminants
atmosphériques dans le ventilateur, assurez-vous que la réserve d’oxygène reliée au ventilateur est
propre et filtrée de manière adéquate
90
, et que le bouchon de l’orifice d’alimentation en oxygène est
correctement installé à chaque fois que le ventilateur n’est pas relié à une source d'oxygène externe.
90
In addition to the existing internal O
2
Inlet filter, P/N 14313, an External, In-Line Oxygen Filter (P/N 14470) is
available from Pulmonetic Systems, Inc.