Page 6-12 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
Low Pressure
The Low Pressure alarm can be set to apply to All breaths or to Volume Control and Pressure
Control breaths only. (For information on selecting breath types, see Chapter 10 - Extended Features,
Low Peak Pressure Alarm.) The Low Pressure alarm establishes the minimum expected circuit
pressure for the selected breath types. If the circuit pressure does not meet or exceed the Low
Pressure setting:
• A LOW PRES alarm is displayed
• The audible alarm is sounded
The Low Pressure alarm is not active in NPPV mode.
To set the Low Pressure alarm:
1) Push the Low Pressure button.
2) Change the setting using the Set Value knob.
Range: “- -”, 1 - 60 cmH
2
O
WARNING
Patient Circuit Accessories - The use of accessories such as Speaking Valves, Heat-Moisture
Exchangers and Filters create additional patient circuit resistance and in the event of a disconnection,
may impede the generation of a Low Pressure Alarm. Ensure that the Low Pressure Alarm settings
accommodate these types of accessories when used in combination with patient circuits.
AVERTISSEMENT
Accessoires du circuit du patient - L’utilisation d’accessoires tels que les membranes vocales, les
échangeurs thermohydriques et les filtres, produit une résistance additionnelle dans le circuit de
patient et en cas de débranchement, elle risque d’empêcher la génération de l’alarme de basse
pression. S’assurer que les paramètres de l’alarme de basse pression s’adaptent à ces types
d’accessoires lorsqu’ils sont utilisés avec les circuits du patient.