Operator’s Manual LTV
®
Series ventilator Page 15-13
p/n 10664, Rev. Y
Symptoms Possible Causes What to Do
Patient circuit leaks.
Ventilator must run
harder to maintain
PEEP.
Perform a Leak Test and reseat or replace the
leaking parts or connections. See page 11-8 for
instructions.
Ventilator gets
excessively hot.
Internal problem with
the ventilator.
Immediately contact a certified Pulmonetic
Systems’ service technician.
Communications
setting is not set to
MONITOR mode.
Set communications setting to MONITOR mode.
See page 10-16 for instructions.
Ventilator requires
upgrades to be
compatible with LTM
Graphics Monitor.
Check LTM compatibility in the Model Number
menu. See page 10-19 for instructions.
If the ventilator is not LTM compatible, it will
require upgrading by a certified Pulmonetic
Systems’ service technician to accommodate the
LTM Graphics Monitor.
Ventilator does not
work with LTM
Graphics Monitor.
Defective connections
between the LTM
Graphics Monitor and
the ventilator.
Check the Communications Data Cable
connection between the ventilator’s
communications port and the LTM Graphics
Monitor’s Data Port. See the LTM Graphics
Monitor Operator’s Manual, P/N 11010, for
detailed instructions.