Operator’s Manual LTV
®
Series ventilator Page C-21
p/n 10664, Rev. Y
Procedure:
Set the O
2
% control to 22%
(LTV
®
1000 Only)
C-18
2) B)
LOW O2 PRES alarm activates after a short
pause
Reset O
2
% to 21 and clear the
alarm. Set the Low Min Vol Alarm to
10 L
C-18
2) C)
LOW MIN VOL alarm activates
Reset the Low Min Vol Alarm to 1.0
and clear the alarm.
Set the Low Pressure alarm to 60.
C-18
2) D)
LOW PRES alarm activates
Set the Low Pressure Alarm to 5
and clear the alarm.
Set the High Pres Limit to 10
cmH
2
O below the Peak Inspiratory
Pressure.
C-18
2) E)
HIGH PRES alarm activates.
Reset the High Pressure Limit alarm
to 100 and clear the alarm.
C-18
2) F)
Disconnect the high pressure sense
line from the ventilator
C-19
2) G)
DISC/SENSE alarm activates on the next
breath
Reconnect the high pressure sense
line and clear the alarm
C-19
2) H)
Selected Monitors should read as follows:
PIP: 36 to 44 cmH2O
PEEP: 17 to 23 cmH2O
Change control settings as follows:
Mode: Pressure, Assist/Cntl
Pressure Control: 40
PEEP: Max
(LTV
®
950 / 1000 Only)
C-19
2) I)
No alarms activate
POWER LOST alarm activates
Battery Level LED illuminates showing the
charge level.
Disconnect AC Adapter from
Ventilator at pigtail cable connector.
C-19
2) J)
Ventilator continues to operate from the
internal battery