Page E-2 LTV
®
Series ventilator Operator’s Manual
p/n 10664, Rev. Y
For software version 3.11, a quantity of 1 is displayed in the initial occurrence
recordings and a quantity of 2 or more in the secondary occurrence recordings of the
same type of event.
5) Push the Select button.
xx:eventdate is displayed.
xx is the chronological number of the event occurrence.
eventdate is the date
98
of the first occurrence (for software versions 3.01 or earlier);
For software version 3.11, the date of the first occurrence is displayed in the initial
occurrence recordings and the date of the latest occurrence in the secondary
occurrence recordings of the same type of event.
6) Push the Select button.
xx:hh:mm:ss is displayed.
xx is the chronological number of the event occurrence.
hh:mm:ss is the time of the first occurrence (for software versions 3.01 or earlier);
For software version 3.11, the time of the first occurrence is displayed in the initial
occurrence recordings and the time of the latest occurrence in the secondary
occurrence recordings of the same type of event.
7) Push the Select button.
xx:data is displayed.
xx is the chronological number of the event occurrence.
data is the data associated with the first occurrence of this event (for software versions 3.01
or earlier);
For software version 3.11, the data associated with the first occurrence is displayed in
the initial occurrence recordings and the data associated with the latest occurrence in
the secondary occurrence recordings of the same type of event.
For some events, the data field will be blank.
8) Push the Select button to return to the initial display.
9) Turn the Set Value knob clockwise or counterclockwise to view other events.
10) To exit the EVENT TRACE, turn to EXIT and push the Select button or push Control Lock.
For more information about how these codes are used, see the LTV
®
Series ventilator Service
Manual (P/N 10665) or contact a certified Pulmonetic Systems’ service technician.
98
Date is displayed in the currently selected date format.