Operator’s Manual LTV
®
Series ventilator Page 6-7
p/n 10664, Rev. Y
NOTE
The ventilator will not perform an Inspiratory Hold maneuver during Pressure Control, Pressure
Support or Spontaneous breaths.
If the button is held during exhalation or any non-volume inspiration:
The associated LED will be blinking.
All buttons that are not lockable will operate normally.
All buttons that are lockable will be ignored.
If the button is released before the inspiration is complete, the display will
return to INSP HOLD.
Once the maneuver is completed, if any buttons are touched or an alarm
occurs, the Δ Pres, C Static or P Plat display will be cleared.
After 60 seconds, the display will be cleared.
REMARQUE
Le ventilateur n’effectue pas une manœuvre de maintien de l’inspiration au cours du contrôle de
pression, du soutien de pression et de ventilations spontanées.
Lorsque le bouton est maintenu pendant l’exhalation ou toute inspiration sans volume:
Le LED correspondant demeure allumé en continu.
Tous les boutons non verrouillables fonctionnent normalement.
Tous les boutons verrouillables sont ignorés.
Lorsque le bouton est relâché avant la fin de l’inspiration, l’affichage
indique INSP HOLD.
Au terme de la manœuvre, l’actionnement d’un bouton ou le déclenchement d’une alarme entraînent
l’effacement de l’affichage Δ Pres ou C Static.
L’affichage est effacée après 60 secondes.