Remote Commands Reference
R&S
®
RTC1000
212User Manual 1335.7352.02 ─ 02
DISPlay:INTensity:BACKlight <Intensity>
Defines the intensity of the background lighting of the display. *RST does not change
the intensity.
Parameters:
<Intensity> Value in percent
Range: 10 to 100
DISPlay:INTensity:GRID <Intensity>
Defines the intensity of the grid on the screen. *RST does not change the intensity.
Parameters:
<Intensity> Value in percent
Range: 0 to 100
DISPlay:PERSistence:STATe
<State>
Defines whether the waveform persists on the screen or whether the screen is
refreshed continuously.
Parameters:
<State> ON | OFF
ON
The waveform persists for the time defined using DISPlay:
PERSistence:TIME.
OFF
The waveform does not persist on the screen. Only the currently
measured values are displayed at any time.
*RST: OFF
DISPlay:PERSistence:TIME <Time>
Persistence time if persistence is active (see DISPlay:PERSistence:STATe).
Each new data point in the diagram area remains on the screen for the duration
defined here. To set infinite persistence, use DISPlay:PERSistence:INFinite.
Parameters:
<Time> Range: 50E-3 to infinite
*RST: 50E-3
Default unit: s
DISPlay:PERSistence:INFinite <InfPersistence>
Sets the persistence time to infinite if DISPlay:PERSistence:STATe is ON. Each
new data point remains on the screen infinitely until this setting is changed or the per-
sistence is cleared.
Display

Table of Contents

Other manuals for R&S RTC1000

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S RTC1000 and is the answer not in the manual?

R&S RTC1000 Specifications

General IconGeneral
BrandR&S
ModelRTC1000
CategoryTest Equipment
LanguageEnglish

Related product manuals