Frozen Tissue Section Preparation Instrument
Cryostat Microtome

Table of Contents

Questions and Answers

  • J
    Jose HawkinsAug 20, 2025
    What to do if nothing is displayed on the screen of Sakura Medical Equipment?
    • J
      Jamie HodgesAug 20, 2025
      If nothing is displayed on the screen of your Sakura Medical Equipment, try the following: * Turn on the power. * Connect the power cable or plug. * Check the breaker on the facility side. * If the instrument power was temporarily cut off, turn the instrument power off and then back on. * Touch the screen.
  • A
    Anthony JacksonAug 25, 2025
    Why has the vacuum suction force dropped in my Sakura Medical Equipment?
    • D
      dominiquecontrerasAug 25, 2025
      If the vacuum suction force has dropped in your Sakura Medical Equipment, check these: * Remove any contaminants trapped in the vacuum suction port using tweezers. * Replace the vacuum filter set if contaminants are trapped on the vacuum filter. * Remove any contaminants trapped in the hose. * Securely connect the vacuum filter and hose if the connection is loose. * Raise the set value of desired suction force if it is set too low.
  • J
    John VaughnAug 26, 2025
    How to unlock the handwheel on Sakura Medical Equipment?
    • K
      Kelsey WeaverAug 26, 2025
      If the handwheel of your Sakura Medical Equipment does not move, check if the handwheel lock tab is engaged and disengage it. Also, if motorized sectioning is paused, press the stop key to cancel the pause.
  • S
    Shawn BruceAug 28, 2025
    What to do if the foot switch does not respond on Sakura Medical Equipment?
    • B
      Brian CarterAug 28, 2025
      If the foot switch of your Sakura Medical Equipment does not respond, check the following: * Disengage the handwheel lock tab. * Unlock the foot switch. * Connect the cable. * Select the MOTORIZED Tab.
  • J
    jenny81Aug 29, 2025
    Why motorized sectioning (right panel) cannot be performed on Sakura Medical Equipment?
    • E
      Eric HowardAug 29, 2025
      If motorized sectioning cannot be performed on your Sakura Medical Equipment, ensure that the MOTORIZED Tab is selected and that the handwheel lock tab is disengaged.
  • L
    Louis BondAug 31, 2025
    Why the chamber cannot be cooled in Sakura Medical Equipment?
    • K
      Kyle DanielAug 31, 2025
      If the chamber of your Sakura Medical Equipment cannot be cooled, readjust the setpoint temperature if it is set too high. Also, check if the operating environment temperature is appropriate.
  • I
    ilandrySep 2, 2025
    What to do if the object holder cannot be secured in the specimen head of the specimen holder of Sakura Medical Equipment?
    • J
      John BeanSep 2, 2025
      If the object holder of your Sakura Medical Equipment cannot be secured in the specimen head, clean the specimen head setting position and the object holder placement surface. If the object holder has gotten smaller due to cleaning with disinfectants, replace it. If the specimen holder lock spring has broken, replace it. In an emergency: 1. Remove the blade holder temporarily. 2. Confirm the handwheel is locked. 3. Raise the specimen head locking lever to place and lock the object holder on the specimen holder. 4. Tighten the nut located at the bottom of the specimen holder using the nut driver provided. 5. Dry the blade holder if it has water droplets on its surfaces before putting it back. 6. Remove the object holder from the specimen holder in the reverse order to locking it.
  • W
    Warren AshleySep 4, 2025
    Why doesn't the blade holder move smoothly to the left/right on Sakura Medical Equipment?
    • R
      Rachel GriffinSep 4, 2025
      If the blade holder does not move smoothly to the left or right on your Sakura Medical Equipment, remove the blade holder from the chamber. Dry it thoroughly and return it to the chamber after it warms to room temperature.
  • G
    Gregory EdwardsSep 6, 2025
    What to do if the window will not open on Sakura Medical Equipment?
    • T
      Tamara YoderSep 6, 2025
      If the window of your Sakura Medical Equipment will not open, wait for the ozone disinfection cycle to complete. If it still does not open after the remaining ozone disinfection time reaches "0", contact the nearest authorized Sakura distributor or representative.
  • T
    Thomas GarnerSep 8, 2025
    How to perform sectioning although the chamber cannot be cooled on Sakura Medical Equipment?
    • M
      Mary MartinSep 8, 2025
      If you want to perform sectioning although the chamber of your Sakura Medical Equipment cannot be cooled, use the emergency plastic holder (optional) to suppress heat conduction from the specimen head, which makes it difficult for the specimen to melt.