USER MANUAL (Instruction for Use) SW 6.1
Treatment Selection
Version 6.1.6 dated 2020-01-31 EN Page 147 of 201
the treatment, the live image of the video camera is displayed if this option is installed on the
AMARIS system.
CAUTION!
The treatment progress diagram displays the percentage of the already
performed treatment. If the laser interrupts the treatment without displaying
100% performed treatment, there is an indication of system failure.
Please contact the Service Department of SCHWIND eye-tech-solutions or its
local distributor and export the treatment log files.
At the bottom of the treatment menu the possibility exists to set and start a timer by pressing the
<Start timer> button and to deactivate the active eye tracking by pressing <Eye tracker>.
IMPORTANT NOTE
The <Timer> button only exists if the timer was activated for this step in the
Treatment Assistant Manager (TAM).
The input of a treatment offset is only possible for Aberration-Free treatments
in the step Treatment.
There is also a section for the optional OCP (Online Coherent Pachymetry) system. By pressing
the <Get Pachy> button, the corneal thickness values can be stored for every particular step of
the procedure.
IMPORTANT NOTE
The functions of the Online Pachymetry can only be used if the Online
Pachymeter is integrated into the AMARIS excimer laser.
On the right side of the menu, a treatment assistant is displayed which guides the surgeon
through the surgical procedure. The surgical steps, automatic hardware adjustments and
messages displayed on the screen can be edited in advance in the Treatment Assistant Manager
(please refer to chapter 7.5 Treatment Assistant Manager).
If the option in the Treatment Assistant Manager is set, the assistant will automatically jump to
the next step. Otherwise, it is possible to proceed manually to the next step by pressing the
<Next> button.
On the upper right side of the Menu there is a button <Surgery comment> with which free text
comments can be input. The comment entered in the treatment screen will be shown on the
treatment printout together with the comments which can also be entered in the SCHWIND CAM
software.
During the TAM step treatment the button <Surgery comment> will be disabled and it is not
possible to enter comments during this step.