IMPORTANT NOTE
Never remove the storage medium (SD-Card) while data is transferred or
during treatment. This may lead to irreparable loss of data. For safety reasons,
removal or input of storage media may only be performed in the Main menu.
8.4 Recovery Function
In case a treatment was aborted by a technical defect or other reason it can be recovered and
continued with the help of the Recovery Function.
The button <Recovery> will appear as active as soon as files for a recovery are available. By
pressing the button the recovery menu will be opened.
It appears a list of available treatments for recovery and the user can select the treatment to be
continued. The maximum number of treatments for recovery is 10.
After the treatment was selected and the button <Continue Selected Treatment> was pressed,
the application software will start the recovery treatment.
IMPORTANT NOTE
A Recovery treatment is valid for 60h after the abort of the original treatment.
After this time it will be deleted automatically from the recovery list.
For a successful recovery of a treatment the actual Fluence value of the
AMARIS has to be within +/- 5% of the Fluence level with which the original
treatment was performed.
If the original treatment was aborted at less than 5% or more than 95% of
treatment progress a message will pop up:
In case the initial treat has used a treatment offset it will also be applicated for
the recovery. A new input of a treatment offset will not be possible then.