5.2 Room and Installation Requirements
IMPORTANT NOTE
The detailed room and installation requirements for SCHWIND AMARIS
excimer laser systems and further technical requirements are listed in the
document: “Room and Installation Requirements for SCHWIND AMARIS
Excimer Lasers 750S / 500E / 1050RS”, which is provided by SCHWIND eye-
tech-solutions.
5.2.1 General Requirements
1. Avoid direct sunlight, use darkening curtain if necessary.
2. Air conditioning required.
3. Air dehumidifier is absolutely necessary in countries with a humidity of more than 40 %.
4. Atmospheric pressure: 860 hPa to 1060 hPa
5. On the outer side of the entrance door, a laser warning lamp and the official laser-warning
symbol must be installed.
6. A switch, which interrupts the laser if the door is opened, can be connected through the
interlock input at the system. This is only recommended, however, when it is guaranteed
that the door will not be opened during treatment.
7. According to EN 207, protective glasses have to be available in sufficient quantity near the
entrance door.
8. Door width: min. 80 cm.
9. Floor covering: PVC or stable, vibration-free underground! (No carpet).
IMPORTANT NOTE
Good room lighting is important in the area where laser eye protection is
worn. Bright diffused wall surfaces are thereby helpful.
For detailed information reg. the “Vibration characteristics of the floor/ground” please refer to a
SCHWIND document “Room and Installation Requirements for AMARIS Excimer Lasers 750S /
500E / 1050RS”.