9 CLEANING AND MAINTENANCE
9.1 General Notes
WARNING!
Damage of device! Danger of short circuit!
Power must always be OFF during cleaning, disinfection or sterilization the
excimer laser, its components or control elements!
Disconnect the patient bed from power supply before and during manual
cleaning and disinfection!
Cleaning and disinfection when device is not switched off can lead to damage
of the device and / or to personal injury by a short circuit.
CAUTION
Energy impairment of the laser beam! Damage of device!
To clean the SCHWIND AMARIS, its parts and components, the SCHWIND
Patient Bed and the treatment room do not use any liquids containing
ammoniac or alcohol.
Do not disinfect your hands near the excimer laser.
Before applying disinfection liquids, please check the ingredients of the agent
in order to exclude that alcohol or ammonia is included.
If necessary contact the manufacturer of the liquid.
Ammoniac and alcohol reduce the energy of the laser beam and can cause
varying of treatment results.
CAUTION
Damage of device!
Do not use ether, acetone or aggressive detergents on painted surfaces! They
can damage the surfaces of the device or damage the medical device itself.
Apply only cleaning and disinfection solutions that are intended for medical
instruments.
Pay attention to expiration date of cleaning and disinfection solutions.
After opening, use solutions within the time frame recommended by the
manufacturer.
Do not use solutions with visible contamination.
Do not use abrasives, scrapers or brushes for cleaning! This can lead to
reduced precision of the parts and to unforeseen damage.