6.4.4 Disconnection from the Electronic Circuit
Complete separation from the electronic circuit is only possible through:
Disconnection of the net connection plug at the AMARIS device,
Or
Switching off of the main switch at the AMARIS (refer to chapter 6.4.1 Main Switch).
6.5 UPS - Mains Failure or Power Breakdown
Power loss of mains
Power loss of mains is indicated by the repeated sound of a beeper of the UPS. In this case the
UPS will automatically resume the power supply of AMARIS for approximately 10 minutes, for
both laser treatment mode and standby mode (no substantial difference in power consumption).
This is enough time to finalize an ablation already in progress. No further ablations shall be
started.
The user is requested to stop all activities concerning the AMARIS within 10 minutes and to
completely shut down the AMARIS.
IMPORTANT NOTE
In no case should a new laser treatment be started later than 2 minutes after
mains power loss.
Please be aware that autonomy of the UPS depends on age and battery condition.
Return of mains power
When mains power returns after failure the user must check the charge condition of the UPS. It
is recommended only to start a new laser treatment if charge condition of the UPS is more than
80%.
IMPORTANT NOTE
Charging the UPS batteries takes up to 24 hours.
WARNING!
Risk of incomplete laser treatment!
There is a risk of incomplete laser treatment if charge condition of UPS is less
than 80% and mains power fails.
Operation of the UPS
For operation of the UPS refer to the USER Manual for the respective UPS type.
Treatment aborted
In case a treatment was aborted refer to 8.4 Recovery Function.