5 INSTALLATION
5.1 General Installation Notes
There are no activities by the user necessary regarding the installation of the medical device or
software.
Trained service personnel of the SCHWIND eye-tech-solutions or of an authorized distributor
perform the installation of the SCHWIND AMARIS and the SCHWIND patient bed.
IMPORTANT NOTE
Please contact your authorized local SCHWIND representative or the Service
Department of SCHWIND eye-tech-solutions to organize the installation of the
software and/or device.
WARNING!
Risk of injury! Damage of device!
Do not install and operate the SCHWIND AMARIS in rooms, areas where
danger of explosion exists and in the presence of flammable mixtures!
Improperly installed cables and lines can smoulder or catch fire.
Live cable ends and components can cause electric shock. To prevent the user
or patient from being injured, make sure that all cables are laid in such a way
that they are protected against damage and cannot be folded or squeezed.
Do not use portable multiple socket outlets sharing a common grounding cable
to connect several devices together with this device to mains.
Do not use multiple cables or any other extension cables. Only the main
connection cable delivered by SCHWIND eye-tech-solutions should be used.
Otherwise the leakage current of the system could be potentially hazardous to
patients!
Misplaced or improperly mounted parts may fall and cause personal injury.
CAUTION
Damage of device!
The SCHWIND AMARIS should not be installed in damp rooms. Avoid
dripping, standing, or splashing water near the device.
The device may only be installed and operated in rooms with a compatible
electrical supply and in which the electrical installation complies with the
relevant national regulations.
Electrostatic processes can damage electronic assemblies.
Consider also the Restrictions of Use and Safety Precautions listed in chapter 2.3 and the
information contained in chapter 5.5 Initial Installation / Start-Up.