USER MANUAL (Instruction for Use) SW 6.1
Maintenance
Version 6.1.6 dated 2020-01-31 EN Page 180 of 201
The AMARIS Application software has an integrated test which checks if the current particle
aspiration filter is older than 28 days. If this is the case a message will appear at login to remind
the user to exchange the particle aspiration nozzle (see Figure 10-2). This is confirmed by entering
the serial number of the new particle aspiration filter into the AMARIS Application software.
Change of the Nozzle for Particle Aspiration System
To exchange the debris removal nozzle enter the LOT number of the new nozzle (see Figure
10-3).
Then start a fluence test, bring the debris removal in position and exchange the old versus
the new nozzle.
There is a security clamp which has to be released before it is possible to pull the old nozzle
out (see Figure 10-4).
Figure 10-2: Reminder for exchange of nozzle for particle aspiration system