the system does not request an update of the scanner positions, and the scanners remain on the
previously calculated position for this pulse.
IMPORTANT NOTE
The 7D tracking functionality will effect all other 6 dimensions in form of a
real time offset.
8.6 Online Coherence Pachymetry (OCP)
For the measurement of the corneal thickness during the treatment, an Optical Coherence
Pachymeter is optionally integrated in the AMARIS excimer laser.
Figure 8-23: Online Coherence Pachymeter
The section for displaying the measured corneal thickness and the raw signal from the OCP in
the treatment screen of the AMARIS Application software is on the right side of the screen just
below the live video image.
The values are updated with a frequency of 1 Hz and a summary of all measured values during
the treatment appear on the treatment printout in the form of a diagram.