USER MANUAL (Instruction for Use) SW 6.1
Installation
Version 6.1.6 dated 2020-01-31 EN Page 84 of 201
5.2.4 Environmental Conditions
Air condition:
Strictly requested!
Heat load: 1.5 kWh
Contaminants: Minimize dust; No smoking; No volatile cleaning, agents/
detergents (Alcohol, Keton, etc.); No aromatic substances;
IMPORTANT NOTE
The conditions of the operating room lies under the direct responsibility of the
user / operator and his quality management system.
The operating conditions shall be established according to standards which are
relevant in the respective country.
Room Temp. and Humidity - see Pos. 1 and 2
1. Operation Mode and Standby
Room Temperature Range: 21°C (70°F) ideal (range 18°C (64°F) - 24°C (75°F)),
Stability ± 2°C within 2h before starting operation and between
fluence test interval.
Relative Humidity:
40 % ideal (range 30% - 45%)
Stability ± 5% rH within 2h before starting operation and between
fluence test interval.
If NO, an air dehumidifier is strictly requested!
Absolute altitude of the
installation location (height
above sea level):
Max. 3000m (due to electrical safety)
Default Setting: 0 m to 250 m
Pressure sensor for the vacuum beam path has to be adjusted to
the actual altitude by the service technician during installation.
2. OFF-mode
Room Temperature Range: 10°C – 30 °C or 50°F – 86°F
Avoid extreme temperature changes.
Relative Humidity:
15% - 85%
NONCONDENSING! Avoid extreme temperature changes!
See User Information UI_A02-09 “Influence of humidity and
condensation on SCHWIND products”.