USER MANUAL (Instruction for Use) SW 6.1
Maintenance
Version 6.1.6 dated 2020-01-31 EN Page 182 of 201
WARNING!
Risk of injury!
Do not attempt to change the nozzle if a patient is situated on the patient
bed!
IMPORTANT NOTE
The used Particle Aspiration filter has to be disposed in a professional way
only (medical special refuse).
10.4 Maintenance of the Plume Evacuation System
The AMARIS can be equipped either with the plume evacuation system or the particle aspiration
system. The plume evacuation system can be recognized by a nozzle with one channel on the
right side (see the left Figure 10-5).
When replacing the nozzle, it must be ensured that the same type of nozzle is available. Check
the type of nozzle before unpacking. See the following images for illustration.
Figure 10-5: Replacing a single-channel nozzle
If the new nozzle you want to install is not of a similar type as the previous one, please contact
the SCHWIND hotline or your local distributor.
IMPORTANT NOTE
The change Plume Evacuation System nozzle should be performed once a
year by a service technician only, tipically, during the annual service visit.