3. Transport and Storage
Vacuum packaging in wooden box for outdoor transport.
Important: If transport without vacuum packaging, the environmental conditions for OFF-mode
(2.) are required.
Environmental
Temperature Range:
0°C – 50 °C or 32°F – 122°F
Avoid extreme temperature changes.
Relative Humidity:
15% - 95%
NONCONDENSING! Avoid extreme temperature changes!
See User Information UI_A02-09 “Influence of humidity and
condensation on SCHWIND products”.
IMPORTANT NOTE
If a change in humidity of more than 5% or a change in temperature of more
than ±2°C is recognized in a shorter time period we recommend performing
an additional fluence test.
5.2.5 Installation Requirements
Input Requirements
For the required performance data of power supply for SCHWIND AMARIS, please refer to the
data in chapter 11 Technical Data.