148 FLOWSIC100 · Operating Instructions · 8012513/YSA5/V2-1/2016-07 · © SICK Engineering GmbH
Start-up and Parameter Settings
Subject to change without notice
Assigning a new IP address using SOPAS ET
Connect the device
p. 116, § 4.1.3.
NOTICE:
Malfunctions in data transfers not caused by the measuring system can
occur during communication via Ethernet.
The FLOWSIC100 manufacturer assumes no responsibility for
malfunctions that may occur during equipment operation when
measured value transfers and their usage to control processes run solely
via Ethernet.
Increasing the value in the "Scan timeout" field to 3000 ms can minimize
communication problems.