Start-up and Parameter Settings
FLOWSIC100 · Operating Instructions · 8012513/YSA5/V 2-1/2016-07 · © SICK Engineering GmbH 149
Subject to change without notice
4.3.4.1 Changing the field bus address for the Profibus module
The Profibus DP interface module is set to field bus address 126 at the factory. Complete
the following procedure to change the address:
Check in the "Configuration / System Configuration“ directory (
p. 146, Fig. 124) that
the interface module (field "Interface Module") is set to "Profibus DP".
Select the "Configuration / I/O-Configuration / Interface Module“ directory and enter
the new address in the "Fieldbus address" window (field "Profibus DP Configuration”).
Fig. 126 "Configuration / I/O Configuration / Interface Module" directory