Start-up and Parameter Settings
FLOWSIC100 · Operating Instructions · 8012513/YSA5/V 2-1/2016-07 · © SICK Engineering GmbH 147
Subject to change without notice
4.3.4 Configuring the Ethernet module
Assigning the Ethernet module a new IP address
An IP address specified by the customer is entered at the factory when the address is
available when the device is ordered. If not, standard address 192.168.0.10 is set.
Complete the following procedure to change the address:
Select directory "Configuration / I/O-Configuration / Interface Module".
Set the desired network configuration in the "Ethernet Interface Configuration" field and
click "Reset module" under "Expansion module information".
Fig. 125 "Configuration / I/O Configuration / Interface Module" directory
NOTICE:
There is a risk of undesired access to the measuring system during
communication via Ethernet.
Only operate the measuring system behind suitable protection devices
(e.g.: Firewall).
New address