EasyManuals Logo

SICK MCS100E User Manual

SICK MCS100E
142 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #131 background imageLoading...
Page #131 background image
Utility programs and protocols
MCS100E Operating Instructions 8009504/VYWA7/V3-1/2018-01 © SICK AG 131
Subject to change without notice
Start End Data Type Description Remarks
----------------------------------------------------------------------------
Component 16 (= O2 ) Values
0951 0952 U32 Date Stamp last Calibration
0953 0954 U32 Time Stamp last Calibration
0955 0956 Float Actual Concentration PKnz
0957 0958 Float mExt add. and mult. correction PExtCor
0959 0960 Float mExt additive correction PExtAdd
0961 0962 Float mExt with ZERO-correction PExtRawN
0963 0964 Float mExt without ZERO-correction PExtRaw
0965 0966 Float Analog Input Signal [0..100%] [0..4095digit ]
0967 0968 Float Input Signal [mV -250..250]
0969 0970 Float Calib.value [Conc] gas1 nominal
0971 0972 Float first CAL [Conc] gas1
0973 0974 Float last CAL [Conc] gas1
0975 0976 Float actual CAL [Conc] gas1
0977 0978 Float first CAL [Conc] gas2
0979 0980 Float last CAL [Conc] gas2
0981 0982 Float actual CAL [Conc] gas2
0983 0984 Float Calib.value [Conc] gas2 nominal
0985 0986 Float Extra Factor Pfactor
0987 0988 Float Cal.Factor PCalF
0989 0990 Float Cal.Factor internal Standard PCalFc
0991 0992 Float Cal.Value internal Standard PCalVc
0993 0993 U16 Gain Level Reference Signal GiRef
0994 0994 U16 Gain Level Measure Signal GiMes
0995 0996 Float Measure Range 1 RngUp[1]
0997 0998 Float Measure Range 2 RngUp[2]
0999 1000 U16 Reserved
:
: Component 24 Values
1351 1352 U32 Date Stamp last Calibration
1353 1354 U32 Time Stamp last Calibration
1355 1356 Float Actual Concentration PKnz
:
1399 1400 U16 Reserved
Modbus system parameter block
register offset = 1400
Start End Data Type Description Remarks
-------------------------------------------------------------------
1401 1401 U16 DevAddr Device Address
1402 1402 U16 AddrOffset Address Offset
1403 1407 U16 RegOffset[1..5] Register Offsets
1408 1410 U16 CoilOffset[1..3] Coil Offsets
1411 1411 U16 DataLogOffset Register Offset DataLog
1412 1412 U16 RegMode Dataformat Results,Comp.
1413 1413 U16 RegFactor 16 Bit Datafactor
1414 1415 U32 DateOffset Julian Date Offset
1416 1416 U16 DataLogFlagMeasure Flag “MEASURE“
1417 1417 U16 DataLogFlagCalib Flag “CALIBRATION“
1418 1421 U16 Reserve[1..4] Reserve
Datalog block
register offset = 1500
Start End Data Type Description Remarks
-------------------------------------------------------------------
1501 1501 U16 Analogchannel(1..32) read and write
1502 1502 U16 day (1..31) read and write
1503 1503 U16 hour (0..23) read and write
1504 1504 U16 Reserve
1505 1505 U16 1.value hour:00 read only
1506 1506 U16 2.value hour:01 read only
: : : : :
1564 1564 U16 60.value hour:01 read only