Examination
Cassette Exposures in the Spotfilm Device (optional)
22 / 52 AXD3-340.620.13.01.02 Operator Manual
Tomographic sequence
◆ Press the exposure release button up to the first pressure point.
– The tube assembly stand moves to the head-end tomography starting posi-
tion (at the left with the unit horizontal).
◆ Give the breathing command to the patient.
◆ Press the exposure release down fully and keep it pressed until the tube as-
sembly stand stops in the foot-end position.
– During the exposure the tube assembly stand moves towards the foot end
and automatically remains there following the exposure.
Prepare the next tomographic exposure:
◆ Set a new tomographic height.
◆ Press the exposure release button up to the first pressure point.
– The tube assembly stand moves from the foot-end position towards the
head-end tomography starting position (at the left with the unit horizontal).
With the exposure release button pressed, the tube assembly stand moves back
into the starting position.