POLYDOROS SX 65/80
Integrated Control Console
16 / 36 AXD3-340.620.13.01.02 Operator Manual
Automatic exposure system “0-point technique
General 0
0-point technique presupposes fluoroscopy; it is therefore expedient only for
spotfilming.
The following exposure parameters are determined and set automatically for
direct and indirect technique:
Programmed characteristic curves
Several characteristic curves, which you can program depending on the sys-
tem configuration, are available to you for the exposure.
Programmed characteristic curves in the curve stock
Direct technique: Indirect technique:
Exposure kV
Exposure mA
Exposure kV
Exposure mA
Focal spot
No. Type of curve Voltage
C1
C2
C3
C4
C5
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
40.0 kV
41.0 kV
42.0 kV
44.0 kV
46.0 kV
C6
C7
C8
C9
C10
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
48.0 kV
50.0 kV
52.0 kV
55.0 kV
57.0 kV
C11
C12
C13
C14
C15
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
60.0 kV
63.0 kV
66.0 kV
70.0 kV
73.0 kV
C16
C17
C18
C19
C20
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
77.0 kV
81.0 kV
85.0 kV
90.0 kV
96.0 kV
C21
C22
C23
C24
C25
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
Plateau whole EP
102.0 kV
109.0 kV
117.0 kV
125.0 kV
133.0 kV