POLYDOROS SX 65/80
Integrated Control Console
AXIOM Iconos R200 AXD3-340.620.13.01.02 13 / 36
Fluoroscopy
➩ For further information refer the FLUOROSOT Compact Operator Manual
Releasing fluoroscopy
Fluoroscopy is released via the fluoroscopy foot switch.
Selecting the fluoroscopic mode
1
Fluoroscopy with automatic control
With the automatic stage buttons on the unit, select the required characteristic
curve for the automatic dose rate control.
High-contrast fluoroscopy
1
Is a special case of fluoroscopy with automatic control. It is selected with the 2nd
pressure point of a two-stage foot switch. The mode is indicated acoustically.
Fluoroscopic data
The fluoroscopic data of kV, mA, time and dose area product are displayed on the
integrated generator control console.
Display of the kV and mA values
❏ During fluoroscopy:
– the current kV and mA values are indicated continuously
❏ After the end of fluoroscopy:
– the mean value of kV and mA for the examination
Display of the fluoroscopic time
Continuous display of cumulative time in minutes during the examination
Display of the dose area product
Continuous display of the dose area product accumulated in the examination.
Resetting the fluoroscopic data
1
When the examination is finished, record the fluoroscopic data displayed and
◆ actuate the Reset button.
– The displays for the fluoroscopic data and for the dose area product are
reset to zero.
1
depending on the system configuration